Chapter 4
Figure 4-2: Connecting Internal Shutter
Carefully install the grating turret. Refer to
6.
for complete information.
Place the Turret Compartment cover on the IsoPlane. Install and tighten all cover
7.
screws.
4.1.2

Remove an Internal Shutter

When using a PI-MAX detector (or other detector without an external shutter control
connector) and the IsoPlane has an internal shutter, it is necessary to either remove the
shutter or control the shutter via a Princeton Instruments SHC-EXT external shutter control
box if one is available. This is because there is no way for these detectors to keep the shutter
open or to open the shutter during an acquisition.
Perform the following procedure to remove an internal shutter from an IsoPlane:
Using the 9/64" hex wrench, unscrew the four (4) socket head screws that secure the
1.
turret compartment cover to the top of the IsoPlane. Lift the cover and screws off of the
IsoPlane and set them aside.
Carefully remove the grating turret and place it in a safe place. Refer to
2.
Remove a
Internal Shutter
S
HUTTER
!
CAUTION!
Exercise extreme caution when handling the grating turret so as
not to touch the grating surfaces.
Turret, on page 19 for complete information.
!
CAUTION!
Exercise extreme caution when handling the grating turret so as
not to touch the grating surfaces.
S
C
C
HUTTER
ABLE
Section 3.1, Turret
Installation, on page 15