46
6.3.2

Alignment

Since detectors are mounted directly to the mounting plate on an IsoPlane, the rotational
alignment and focusing operations differ from the way these operations are performed for
Acton SpectraPro Series spectrographs.
Perform the following procedure to align the system:
Mount a light source (e.g., a Princeton Instruments Hg and Ne/Ar Dual Switchable light
1.
source,) in front of the entrance slit.
With the IsoPlane properly connected to the computer, turn on the power and wait for
2.
the spectrograph to initialize.
When using an LN-cooled detector, verify that the Dewar has been properly filled with
3.
LN.
With the detector mounted to the spectrograph and connected to the computer, turn
4.
power on and wait for the detector to initialize.
Launch LightField.
5.
Drag the icons for the desired detector and the IsoPlane into the
6.
area.
Configure the spectrograph as follows:
7.
Wait until the detector locks at its default temperature.
8.
Using a 9/64" hex wrench, loosen the four (4) screws at the corners of the detector
9.
mounting plate.
Select the
10.
Spectrometer Alignment
This dialog, illustrated in
the current setup to assist in performing rotational alignment of the array to the
spectrograph's optics. When the
and continuous live data will be displayed as the detector is rotated.
NOTE:
The following information assumes familiarity with the locations
of the mounting plate, Micrometer Compartment, and the locking
set screw. If not, refer to earlier sections of this manual.
When using a mercury source, configure the spectrograph to 500 nm.
When using a broadband source, configure the spectrograph to 0.0 nm.
Align Spectrometer
dialog.
Figure
function from the
Experiment
6-3, describes the changes that LightField will make to
button is clicked, the modifications are made
Begin
Experiment Devices
menu to open the