Teledyne Everywhereyoulook IsoPlane 320A Manual page 41

Table of Contents

Advertisement

Chapter 5
Figure 5-5: Vertical Alignment Adjustment Set Screws
NOTE:
D
EPENDING ON THE DATE
OF MANUFACTURE
VERTICAL ALIGNMENT SCREWS
MAY
AREA
The following tools are required to perform this procedure:
Perform the following procedure to adjust/fine-tune the vertical alignment:
Remove the Micrometer Compartment cover (see
1.
adjustment screws illustrated in
IsoPlane Setup
,
THE TWO
BE LOCATED IN THIS
.
!
CAUTION!
When performing this procedure, be aware of, and adhere to, the
following guidelines:
Each adjustment made to the Cap and Set Screws should
be incremental, no more than approximately 1/8 turn in
either direction.
If the Cap Screw is turned 1/8 turn clockwise, turn the Set
Screw 1/8 turn counterclockwise.
Similarly, if the Cap Screw is turned 1/8 turn
counterclockwise, turn the Set Screw 1/8 turn clockwise.
A series of incremental adjustments is recommended.
At the end of the procedure the Cap and Set Screws should
be tight.
5/32" hex wrench;
3/32" hex wrench.
Figure
Figure
5-5.
V
A
ERTICAL
LIGNMENT
A
S
S
DJUSTMENT
ET
V
A
ERTICAL
LIGNMENT
A
C
S
DJUSTMENT
AP
5-3) to access the height
37
CREW
CREW

Advertisement

Table of Contents
loading

Table of Contents