Teledyne Everywhereyoulook IsoPlane 320A Manual page 58

Table of Contents

Advertisement

54
b.
After completing the rotational alignment, tighten the four mounting plate screws.
7.
Use a 9/64" hex wrench to remove the cover from the Micrometer Compartment on the
8.
IsoPlane.
Use a 3/32" hex wrench to loosen the micrometer locking set screw located in the
9.
micrometer compartment. See
Figure 6-12: Micrometer Locking Set Screw
While continuously acquiring data, adjust the micrometer until:
10.
Once properly aligned, tighten the micrometer locking set screw.
11.
Place the Micrometer Cover on the spectrograph.
12.
Replace and tighten the four (4) cover screws.
13.
Stop data acquisition.
14.
Using the
Experiment Setup|ROIs
One ROI at the top of the array;
One ROI in the middle of the array;
One ROI at the bottom of the array
Configure the data display on the
is selected on the
3d Graph
The
Z Axis Endpoints
The
Save as Default
Click
. While running in
OK
Aligning the large cursor with a prominent peak, rotate the detector until the peak
aligns with the cursor in all three ROIs.
The intensity level has been maximized, and
The FWHM for a selected peak or peaks has been minimized.
tab configure three 10-row ROIs as follows:
Display Layout
tab;
General
on the
tab are
3D Layout
check box is checked.
mode, three stacked graphs should be visible.
Focus
Figure
6-12.
L
S
OCKING
ET
dialog as follows:
and
%X = 0
% Y = 40
S
CREW
;