Using Winspec; Select The Detector - Teledyne Everywhereyoulook IsoPlane 320A Manual

Table of Contents

Advertisement

Chapter 6

6.4 Using WinSpec

After verifying that the detector and IsoPlane are connected to the computer and are both
turned on, Launch WinSpec. WinSpec should automatically identify the detector that is to
be used, and a dialog for selecting the detector for the experiment should be displayed. In
addition, WinSpec must be instructed/configured to use IsoPlane.
6.4.1

Select the Detector

Perform the following procedure to select the desired detector:
If this is the first time you have installed WinSpec/32 and a supported detector, the
1.
Camera Detection Wizard
detector type, click the
CCD
On the
2.
Figure
Figure 6-6: Typical Camera Detection Wizard: Welcome Dialog
Follow the instructions on the dialogs to complete the initial hardware setup. The
3.
wizard enters default parameters on the
opportunity to acquire a test image to verify that the system is working properly.
When using an LN-cooled detector, fill the Dewar while the detector is powered on.
4.
Operation
NOTE:
When using an LN-cooled detector, the Dewar should be filled
while the detector is ON.
will run automatically. Otherwise, if you installing a new
Launch Camera Detection Wizard...
tab to launch the wizard.
dialog, leave the checkbox unselected and click
Welcome
6-6.
button on the
dialog tabs and provides an
Hardware Setup
49
Controller/
. See
Next >