Focus The Spectrometer - Teledyne Everywhereyoulook IsoPlane 320A Manual

Table of Contents

Advertisement

34
Launch the application software.
4.
Loosen the rotation adjustment screws (see
5.
the detector. The mounting plate holes are slotted to allow approximately 4° of
rotational positioning.
If background subtraction is an available function, acquire a background to be applied
6.
while data are acquired.
Turn on the light source and begin continuous data acquisition.
7.
While viewing the data being acquired, rotate the detector until you obtain the sharpest
8.
peaks or, if in imaging mode, the spectral lines are perpendicular in the image.
Tighten the mounting screws.
9.
Confirm that the data still show sharp peaks or perpendicular spectral lines.
10.

5.5 Focus the Spectrometer

The basic actions for focusing the spectrometer are listed below. This general procedure
assumes that you have already mounted the array detector to the exit port of the
spectrograph and that there is a light source at the entrance port. Refer to Chapter 3 of this
manual for detailed LightField and WinSpec instructions for performing this function. For
other application software, refer to that software's manual or help file.
Using a 9/64" hex wrench, unscrew the four (4) screws that secure the Micrometer
1.
Compartment cover. See
Figure 5-3: Micrometer Compartment Location
Set the screws aside.
2.
Figure
5-3.
M
ICROMETER
C
OMPARTMENT
Figure
5-2) sufficiently that you can rotate