Chapter 6
Figure 6-2: Typical LightField Experiment Settings Stack and Experiment Devices
Note that the
3.
Because this is a new experiment, the default settings for the detector will be active. The
Status
Temperature status reports the current temperature and whether the set temperature has
been reached. Clicking on the icon, opens the
temperature can be changed.
On the
4.
Open the Spectrometer expander and select the appropriate grating. See
5.
this case, the 300 g/mm, Blaze: 750 grating has been selected and the center wavelength
has been set to 500 nm for a Mercury lamp. When using a boradband source, use
0.0 nm.
Operation
Experiment Settings
bar at the bottom of the window displays an icon for temperature status.
expander, select
Shutter
TIP:
Overhead fluorescent lights produce a mercury spectrum. Use a
white card tilted at 45 degrees in front of the entrance slit to
reflect overhead light into the spectrograph. Select 500 nm as the
spectral line.
stack on the left now displays several expanders.
expander which is where the set
Sensor
. See
Shutter Mode: Normal
45
Figure
6-2.
Figure
6-2. In