Figure 5-6. Sample Trend Data Printout - Covidien Nellcor Operator's Manual

Bedside spo2 patient monitoring system
Hide thumbs Also See for Nellcor:
Table of Contents

Advertisement

1
Product column headings
2
Patient data column headings
3
Time column
4
Output Complete
5
%SpO
6
PR
7
PA
8
Status
Ensure patient data settings coincide with expected settings. This would include
2.
the version of firmware and its CRC code, which should be all zeros; the current
method of viewing the data: waveform, trend, or graph; alarm limit settings;
patient mode; and SatSeconds setting.
Scan the time, SpO
3.
Match the operating status codes to the following table for pertinent system infor-
4.
mation. Reference
Operator's Manual
Figure 5-6.  Sample Trend Data Printout
Data source, firmware version, and system settings
Lists appropriate time and data headings
Real-time clock date and time stamp
Message indicating completion of trend data download
Current saturation value
2
Current pulse rate
Current pulse ampltitude
Operating status of the monitoring system
, or PR column until reaching the events of interest.
2
Operating Status
Codes, p. 5-54.