Product Specifications
Sensor and Cable Compliance
11.9.2
WARNING:
The use of accessories, sensors, and cables other than those specified may
result in inaccurate readings of the monitoring system and increased emission
of the monitoring system.
Pulse oximetry sensor cable
Power cable
Nurse call cable
Pulse oximetry interface cable
Safety Tests
11.9.3
Ground Integrity
100 milliohms or less
Leakage Current
The following tables display the maximum earth and enclosure leakage current
allowed, as well as patient leakage.
100
Table 11-9. Cables and Sensors
Item
Sensors
Cables
Maximum Length
1.6 ft. (0.5 m)
10.0 ft. (3.0 m)
5.9 ft. (1.8 m)
10.0 ft. (3.0 m)
Operator's Manual