Preventive Maintenance

Periodic Safety Checks

7.5
Covidien recommends a qualified service technician perform the following
checks every 24 months.
Inspect the equipment for mechanical and functional damage or deterioration.
Inspect the safety relevant labels for legibility. Contact Covidien or a local Covidien
representative, if labels are damaged or illegible.
Ensure all user interface keys, cables, and accessories function normally.

Service

7.6
WARNING:
Only a qualified service technician should remove the cover or access any
internal components.
Caution:
Dispose of monitoring system in accordance with local requirements and
regulations.
The monitoring system requires no routine service other than cleaning, battery
maintenance, and service activity mandated by the institution. For more infor-
mation, reference the Service Manual.
The monitoring system requires no calibration.
Have a qualified service technician replace the battery at least every two (2) years.
If service is necessary, contact Technical Services or a qualified service technician.