SatSeconds™ Alarm Management Feature
The SatSeconds Safety Net
10.6.4
The SatSeconds "Safety Net" is for patients with saturation levels frequently
below the limit, but not staying below the limit long enough for the SatSec-
onds time setting to be reached. When three or more limit violations occur
within 60 seconds, an alarm sounds even if the SatSeconds time setting has
not been reached.
Operator's Manual