Figure 3-1. Connecting A Pulse Oximetry Sensor To Interface Cable - Covidien Nellcor Operator's Manual

Bedside spo2 patient monitoring system
Hide thumbs Also See for Nellcor:
Table of Contents

Advertisement

WARNING:
Use only the Covidien-approved pulse oximetry sensor and interface cables.
Use of another cables can have an adverse effect on performance. Do not
attach any cable intended for computer use to the sensor port.
WARNING:
Failure to cover the applied pulse oximetry sensor with opaque material
while operating under high ambient light conditions may result in inaccurate
measurements.
Caution:
For best product performance and measurement accuracy, use only accessories
supplied or recommended by Covidien. Use accessories according to the
Directions for Use. Use only accessories that have passed the recommended
biocompatibility testing in compliance with ISO10993-1.
Prior to connecting a sensor, perform a safety check of the equipment. Refer-
ence
Periodic Safety
sors, p. 9-77, for details regarding sensor selection.
To fully connect an Nellcor™ pulse oximetry sensor:
Select an appropriate compatible Nellcor™ pulse oximetry sensor for the patient and
1.
desired application. When selecting a sensor, consider the patient's weight and activity,
adequacy of perfusion, availability of sensor sites, need for sterility, and anticipated dura-
tion of monitoring.
Carefully apply the sensor to the patient after reading the Directions for Use
2.
accompanying the sensor. Observe all warnings and cautions in the Directions for
Use.
Connect the interface cable to the sensor port on the front of the panel and firmly
3.
connect the interface cable to the pulse oximetry sensor. When the monitoring
system detects a valid pulse, it enters monitoring mode and displays real-time
patient data.
Operator's Manual
Checks, p. 7-72. Reference
Figure 3-1.  Connecting a Pulse Oximetry Sensor to Interface Cable
Nellcor™ Pulse Oximetry Sen-
Setup
23