Navigation — Rotate the jog dial clockwise or counter-clockwise until a colored
1.
highlight surrounds the desired area. Any rotation of the jog dial either navigates
or changes the desired option setting.
Selection — Press the jog dial to select that desired area, then continue rotating
2.
the jog dial until highlighting the desired menu option, then press again.
The LCD display panel also users with easy-to-read numeric values for patient
oxygen saturation and pulse rate in cyan and yellow, respectively. Reference
Table 2-1.  

Menu Structure

4.4.1
Item
SpO
Menu
2
PULSE RATE (PR) Menu
VOLUME
MODE (Response Mode)
TREND DATA DOWNLOAD
DELETE ALL TREND DATA
SERVICE MENU
SpO
LIMITS options
2
PULSE RATE LIMITS options
Operator's Manual
on page 2-14.
Table 4-1.  Menu structure and available options
QUICK ACCESS ALARM LIMITS Menus
SatSeconds™ alarm management setting (Off, 10, 25, 50, 100)
Upper (21-100) SpO
2
Lower (20-99) SpO
2
Alarm Inhibition for SpO
Upper (30-245) pulse rate alarm limit
Lower (25-240) pulse rate alarm limit
Alarm Inhibition for pulse rate alarms
OPTIONS Menu
Alarm Volume (1-8)
Key Beep Volume (Off, 1-7)
Pulse Volume (Off, 1-7)
Normal, Fast
Start (Cancel or Return), Return
No, Yes
(For qualified service technicians only)
ALARM/LIMITS Menu
Upper (21-100) SpO
2
Lower (20-99) SpO
2
Alarm Inhibition for SpO
Upper (30-245) pulse rate alarm limit
Lower (25-240) pulse rate alarm limit
Alarm Inhibition for pulse rate alarms
Available Selections
alarm limit
alarm limit
alarms
2
alarm limit
alarm limit
alarms
2
Menu Options Navigation
Default
100
Depends on
patient
mode. Ref-
erence
Table 4-4.  
on page 43
5
4
4
Normal
--
--
--
Depends on
patient
mode. Ref-
erence
Table 4-4.  
on page 43.