Oxygen Cylinder Data Screen

4.8.4
The Oxygen Cylinder Data screen is available in the hospital and transport domains and in
standby and ventilating condition.
Estimated cylinder use time can be displayed on the monitor screen if the relevant cylinder data
is entered, the O
Touch More, and then touch O
for the cylinder being used. Finally, touch O
Accept. The ventilator will then calculate the estimated duration of the oxygen supply, and
provide a warning before the cylinder may run out. Allow the calculated display to stabilize for
several minutes before starting a transport, and ensure that the estimated time is sufficient for the
planned trip or outing.
Note:
The oxygen cylinder use time displayed in the monitor is an estimate only. It can be affected by many
factors, such as leaks in the O
remaining pressure level frequently to confirm actual oxygen consumption rate.
Size: Toggle to the size of the oxygen cylinder in use: D, E, H, M, K, 100 L, and 150 L.
Units: Toggle this button to use the desired pressure units for the oxygen cylinder (psi or ATM).
Cylinder Pressure: Touch this button, and use the up/down arrows to enter the current pressure
in the oxygen cylinder (300–2450 psi or 25–175 ATM).
O
Cylinder Monitor: Toggle this button to enabled, and press Accept when all data has been
2
entered.
Operator's Manual
Figure 4-11.  Oxygen Cylinder Data Screen
cylinder monitor is enabled, and the O
2
Cylinder Data Screen. Enter the size, pressure units, and pressure
2
path. Do not rely solely on this measurement. Check the oxygen cylinder
2
monitor on the Utility screen is enabled.
2
Cylinder Monitor to enable the function, and press
2