Table 11-8. Recommended Separation Distances - Covidien Nellcor Operator's Manual

Bedside spo2 patient monitoring system
Hide thumbs Also See for Nellcor:
Table of Contents

Advertisement

Immunity
Test
Conducted RF
IEC/EN 61000-4-6
Radiated RF
IEC/EN 61000-4-3
Rated Maximum
Output Power (P) of
Transmitter in Watts
For transmitters rated at a maximum output power not listed above, estimate
the separation distance (d) using the equation in the corresponding column,
where P is the maximum output [power rating of the transmitter in watts (W)]
according to the transmitter manufacturer.
Note:
Portable and mobile RF communications equipment can affect medical electrical
equipment. Such RF equipment should be used no closer to any part of the monitoring
system, including cables, than the recommended separation distance calculated from
the equation appropriate for the frequency of the transmitter.
Operator's Manual
Table 11-8.  Recommended Separation Distances
IEC/EN 60601-1-2
Test Level
Frequency of
Transmitter
3 Vrms
150 kHz
80 MHz
3 V/m
80 MHz
800 MHz
3 V/m
800 MHz
2.5 GHz
150 kHz to 80 MHz
0.01
0.10
1.00
10.00
100.00
12.00
Compliance
Level
3 Vrms
150 kHz
80 MHz
3 V/m
80 MHz
800 MHz
3 V/m
800 MHz
2.5 GHz
Separation Distance in Meters
80 MHz to 800 MHz
0.12
0.12
0.38
0.38
1.20
1.20
3.80
3.80
12.00
Manufacturer's Declaration
Electromagnetic Environ-
ment Guidance
Equation for Separation
Distance (d)
d
=
1.2 P
150 kHz to 80 MHz
d
=
1.2 P
80 MHz to 800 MHz
d
=
2.3 P
800 MHz to 2.5 GHz
800 MHz to 2.5 GHz
0.23
0.73
2.30
7.30
23.00
99