Installation

Connecting to Power

3.4.1
The monitoring system operates on AC power or on a charged internal battery.
Prior to connecting to power, perform a safety check of the equipment. Refer-
ence
Periodic Safety
To connect the AC power cable:
Ensure the AC outlet is properly grounded and supplies the specified voltage and
1.
frequency (100-240V~ 50-60 Hz).
Connect the female connector end of the AC power cord to the AC power con-
2.
nector on the monitoring system's rear panel.
Plug the male connector end of the AC power cord into a properly grounded AC
3.
outlet.
If necessary, connect grounding wire.
4.
Ensure the Battery Charge Indicator lights.
5.
Note:
Even if the monitoring system is not turned on, the Battery Charge Indicator lights
when the AC power cord is connected into a mains outlet. Reference
Troubleshooting, p. 8-73, if the battery charging indicator does not light when
connected to power.
To troubleshoot an unlit Battery Charge Indicator:
Check the power cord.
1.
Check the AC power inlet.
2.
Check the power/ mains outlet.
3.
Ensure the internal battery is properly installed and charged.
4.
Contact a qualified service technician or a local supplier for assistance.
5.
20
Checks, p. 7-72.
Connect the grounding wire connector to the rear panel's equipotential terminal.
Attach the clip end of the grounding wire to the grounding terminal on the wall.