Second Spo Event; Figure 10-4.   Second Spo 2 Event: No Satseconds Alarm - Covidien Nellcor Operator's Manual

Bedside spo2 patient monitoring system
Hide thumbs Also See for Nellcor:
Table of Contents

Advertisement

Second SpO
10.6.2
Consider the second event. Suppose the SatSeconds alarm limit is still set to
25. The patient's SpO
seconds before saturation again exceeds the lower alarm threshold of 85%.
Because the SatSeconds alarm limit is set to 25 and the actual number of Sat-
Seconds equals 15, there is no audible alarm.
Operator's Manual
Event
2
drops to 84% and the duration of the event is 15
2
1% drop below the lower alarm limit threshold
x15 second duration below the lower threshold
15 SatSeconds; no alarm
Figure 10-4.  Second SpO
SatSeconds™ Alarm Management Feature
Event: No SatSeconds Alarm
2
87

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents