Figure 5-5. Trend Data Download Status - Covidien Nellcor Operator's Manual

Bedside spo2 patient monitoring system
Hide thumbs Also See for Nellcor:
Table of Contents

Advertisement

Data Management
Press the jog dial again, since the item highlighted is the START option.The status
9.
bar indicating total percentage of the download appears and the START option,
immediately changes to a CANCEL option.
Note:
Users may choose to cancel the download operation at any point in the download
process by selecting CANCEL and then RETURN.
Confirm the monitoring system is sending trend data to a personal computer (PC)
10.
by observing the computer screen for a scrolling trend data record. If no trend
data values appear, check connectivity and ensure the personal computer con-
tains HyperTerminal software. If this is all operational, verify patient trend data
history exists on the monitoring system. Contact Technical Services or a qualified
service technician for assistance.
Wait for the OUTPUT COMPLETE message to indicate the download is complete.
11.
Save patient trend data to the personal computer disk or to an alternate source,
12.
depending on institutional requirements.
To launch HyperTerminal
Click the START menu in the main task bar.
1.
Mouse over the PROGRAMS submenu, then ACCESSORIES, then COMMUNICA-
2.
TIONS, then the HYPERTERMINAL option.
52
Figure 5-5.  Trend Data Download Status
Operator's Manual

Hide quick links:

Advertisement

Table of Contents
loading