Covidien Nellcor Operator's Manual page 46

Bedside spo2 patient monitoring system
Hide thumbs Also See for Nellcor:
Table of Contents

Advertisement

Operation
WARNING:
Do not preset different alarm limits for the same or similar equipment within
a single area.
Caregivers may choose to adjust SpO
from default values as necessary. These changes remain in effect until modified
again or until a power cycle occurs. Changes to the SpO
alarm thresholds appear in their respective numerical area. In addition, caregiv-
ers may choose to use the SatSeconds™ alarm option to manage the frequen-
cy of SpO
higher the value, the less frequent the alarm.
SpO
display value flashes zeros during loss-of-pulse alarms and flashes the SpO
on a yellow background when saturation is outside the alarm limits. During pulse
searches, the monitoring system continues to update the display. Current upper
and lower alarm limit settings appear as smaller values to the left of the dynamic
SpO
Pulse Rate (PR) numerical area — Displays the pulse rate in beats per minute
(bpm). The display value flashes zeros during loss-of-pulse alarms and flashes the
pulse rate value on a yellow background when the pulse rate is outside of the
alarm limits. During Pulse Search, the monitoring system continues to update the
display. Pulse rates outside of the pulse rate range of 20 to 250 bpm are displayed
as 0 and 250, respectively. Current upper and lower alarm limit settings appear as
smaller values to the left of the dynamic pulse rate value. Reference
Defaults, p. 4-43, for default alarm limit settings.
To set alarm limits:
Rotate the jog dial to highlight the ALARM LIMITS icon.
1.
Press the jog dial to display the ALARM/LIMITS Menu.
2.
Rotate the jog dial to highlight the desired option.
3.
Press the jog dial to select the desired option.
4.
36
alarm limit violations by adjusting the SatSeconds™ setting. The
2
numerical area — Indicates hemoglobin oxygen saturation levels. The
2
value. Reference
Factory
2
Alarm Limits include pulse rate (PR) and SpO
The SatSeconds™ Alarm option provides alarm management of SpO
limit violations.
The alarm inhibit icon provides caregivers with the option of inhibiting the
alarm for SpO
and/or pulse rate alarms.
2
and pulse rate (PR) alarm thresholds
2
Defaults, p. 4-43, for default alarm limit settings.
alarm limit ranges.
2
and pulse rate (PR)
2
value
2
Factory
alarm
2
Operator's Manual