Pump Parameters

The pump comes programmed with a number of parameters that can be viewed using
'View Set Up' from the main menu and changed using the technician access code in
'Change Set Up' mode.
PARAMETER
Battery Level
Time, Date
Keystroke Volume
Alarm Sound Volume
Priming Volume
Priming Rate
Maximum Bolus
Maximum CA Bolus
Bolus Rate
Maximal Basal Rate
Keep Vein Open Rate
Program Limits
Minimum Lockout
Bolus Titration
Info screen duration
Language
Serial No.
Production Date
Software Version
Main Self-test
Volume Test
Manual Tests
BodyGuard 595 Color Vision™ Directions For Use
PURPOSE
Current battery charge
Set/View time and date
Volume of keystroke
Volume of alarms/alerts
Amount of fluid delivered during
set priming.
Rate at which priming occurs.
Maximum bolus volume for a
protocol
Maximum bolus volume allowed by
a clinician
Rate at which boluses are
delivered
Maximum continuous basal rate
A continuous rate after infusion
program has ended which is
designed to prevent coagulation
and keep an IV line open
MediGuard option
Minimum lockout time between
boluses
Change bolus volume during
infusion
Number of seconds the info screen
is displayed before returning to the
current infusion screen
Change between available
language (may vary between
software version)
Serial Number Unique Pump
Date of manufacture
Pump software version
Automatically test all pumps
hardware features
Test the volume calibration
Manually test the pumps hardware
4 – Operation
RANGE
Empty >> Full
N/A
0-7
4-7
0-20ml
300-1000 ml/hr.
100ml
60ml
40-1200ml
System will adapt rate
according with in-line
backpressure
PCA
0.1 - 100 ml/hr
Epidural 0.1 - 30 ml/hr
Regional 0.1 - 100 ml/hr
0.1-5ml/hr
On/Off
0-240min
ON/OFF
1-255 sec
English, Spanish, Italian,
German (Deutch), French,
Dutch *double check with
Nuray*
Fixed data
Day-month-year
Fixed data
prime volume, continuous
rate, bolus dose, bolus
rate, bolus lock-out
Keypad test, display test,