V102 Failure
V104 Failure
HDF-Pump failure
Float-Switch Failure
Connect Disinfectant
Press CONFIRM key
Please Wait
Disinfectant empty ?
Disinf-Temp. too high Transition to disinfection in Dis I to V.
1-68
Fresenius Medical Care 4008 4/08.03 (TM)
During a cleaning program
in RI I to II, HR I to III, Dis I to V.
V102 has been opened electrically.
24 V are switched off.
The error cannot be acknowledged.
During a cleaning program
in RI I to II, HR I to III, Dis I to V.
V104 has been opened electrically.
24 V are switched off.
The error cannot be acknowledged.
During a cleaning program
in RI I to II, HR I to III, Dis I to V.
The HDF pump has stopped, or the speed deviates (rated value:
400 ml/min, actual value: ≤ 300 ml/min; rated value: 150 ml/min, actual
value: ≤ 100 ml/min).
The error message can be acknowledged for one complete cleaning
program run by pressing the respective program key. The prompt:
"Are you sure ?" is displayed.
During a disinfectant program in the suction phase in Dis I to V (CDS:
Dis I to IV).
The lower switching point of the float switch is not reached within
20 sec.
The program has stopped.
Disinfectant suction phase in Dis V.
Request to connect the disinfectant.
Disinfectant suction phase in Dis V.
After the disinfectant has been connected, the Confirm key on the
menu panel must be pressed to start the suction procedure.
The program has stopped.
Disinfectant suction phase in Dis V.
Disinfectant is drawn in via the concentrate pump.
Disinfectant suction phase in Dis I to V.
Dis V: After the disinfectant has been drawn in, the float switch does not
recognize any fluid.
Dis I to IV, Dis VI: The V84 monitoring unit does not recognize any
conductivity.
Temperature at the end of the rinse-free procedure > 40 °C.
Again and again, the rinse-free procedure is prolonged by 1 minute.
An audible warning is sounded after 4 minutes.
The message is automatically cleared, and it cannot be acknowledged.