Writing/Reading The Digital Outputs Of Cpu I - Fresenius Medical Care 4008 E Technical Manual

Hemodialysis machine
Table of Contents

Advertisement

5.11

Writing/Reading the digital outputs of CPU I

Explanation:
UF Volume display:
UF Rate display:
CPU1: COMBINATION
Confirm
key
CPU 1_COMBI: V24
Valve 24
Activation, P.C.B. LP 634/IC 10/19
Acknowledgement, P.C.B. LP 633/
IC 13/5
CPU 1_COMBI: V24B
Valve 24b
Activation, P.C.B. LP 634/IC 10/13
Acknowledgement, P.C.B. LP 633/
IC 13/4
CPU 1_COMBI: V26
Valve 26
Activation, P.C.B. LP 634/IC 10/18
Acknowledgement, P.C.B. LP 633/
IC 13/3
CPU 1_COMBI: V43
Valve 43
Activation, P.C.B. LP 634/IC 7/15
Acknowledgement, P.C.B. LP 633/
IC 13/2
CPU 1_COMBI: V102
Valve 102
Activation, P.C.B. LP 634/IC 10/12
Acknowledgement, P.C.B. LP 633/
IC 21/4
(Activation of the valve is possible only
if a mandatory rinse is not requested)
5-30
Fresenius Medical Care 4008 4/08.03 (TM)
Acknowledgement/input (in case of 1111, the three status LEDs of the traffic light
are also illuminated)
Activation/output (can be changed with: 4008/E/B:
UF Rate, 4008 H/S: +/–)
CPU 1_COMBI: V104
Valve 104
Activation, P.C.B. LP 634/IC 7/14
Acknowledgement, P.C.B. LP 633/IC
21/5
(Activation of the valve is possible only
if a mandatory rinse is not requested)
CPU 1_COMBI: UF1_PUMP
UF pump 1
Activation, P.C.B. LP 634/IC 4/14
Acknowledgement, P.C.B. LP 633/IC
13/6
(when setting from 0 to 1 = 1 stroke;
acknowledgement is a brief change to
1)
CPU 1_COMBI: AIR_SEP
Switching the ASP on/off
back to menu ?
Confirm
key
CPU 1: COMBINATION