5.7

Writing the analog outputs of CPU I

Explanation:
UF Rate display:
Time Left display:
CPU1: WR ANALOG OUTP
Confirm
key
O: CPU1_TEMP_SET
Rated value specified for the tempera-
ture
IC 53/9
0 to 5 V (X634R/C20, 0.2 to 10 V)
O: CPU1_DAC_DIM
Calibration voltage, dimness
IC 53/8
0 to 5 V (C634R/A11, 0 to 5 V)
O: CPU1_TEMP_ADJ
Calibration voltage, temperature con-
trol
IC 53/7
0 to 5 V (X634R/C21, 0 to 5 V)
O: CPU1_DAC_BLL
Calibration voltage, blood leak
IC 53/6
0 to 5 V
(X634R/A12, 0 to 5 V)
O: CPU1_BIBAG_TEMP_AJ
Calibration voltage for temperature
NTC
(slot X107 / LP 747)
IC 53/5
0 to 5 V (X634R/A13, 0 to 5 V)
DAC value (can be changed with: 4008/E/B:
Analog voltage on P.C.B. LP 634, in 0.1 V
O: CPU1_DAC_X2
Not used
IC 53/4
0 to 5 V (X634R/C13, 0 to 5 V)
O: CPU1_STEUER_EP
Speed setting, degassing pump
IC 53/3
0 to 4.4 V (X634L/ between A, B, C 27
and A, B, C 28, 0 to 21 V)
O: CPU1_STEUER_FP
Speed setting, flow pump
IC 53/2
0 to 4.4 V (X634L/ between A, B, C 29
and A, B, C 30, 0 to 21 V)
back to menu ?
CPU1: WR ANALOG OUTP
Fresenius Medical Care 4008 4/08.03 (TM)
UF Rate, 4008 H/S: +/–)
Confirm
key
5-19