F14 Bloodsystem
F15 Bloodsystem
F16 Bloodsystem
F17 Bloodsystem
1-14
Fresenius Medical Care 4008 4/08.03 (TM)
Applicable for SW 4.91/2.91 and higher if SN, ONLINE-HDF or
4008 HDF pump is connected (= ADKS active)
Acknowledgement that CPU1 detects that the pump is inactive (pump
is not running).
– Acknowledgement (BPSB_VEN,X348V/6) → X633L/A13, 12V
missing
– Control line (BPSST_VEN, X634L/B15) → X348V/1 not 12V or
(BPST_VEN, X634L/A15) → X348V/3 not 12V
– IC16 on P.C.B. LP633 defective
– P.C.B. LP633 recognizes Single-Needle pump although it is not
connected.
Applicable for SW 4.91/2.91 and higher if SN, ONLINE-HDF or
4008 HDF pump is connected (= ADKS active)
CPU1 fails to stop the corresponding blood pump.
CPU2 detects that the pump remains active.
– Control line (BPSST_VEN, X634L/B15) → X348V/1, 0V missing as
well as (BPST_VEN, X634L/A15) → X348V/3 not 0V
– Acknowledgement (BPSB_VEN, X348V/6) → X632/B11, 0V miss-
ing
– Transistor T9 on P.C.B. LP754 defective
– IC5 on P.C.B. LP632 defective
– During the test the lines are inserted on the corresponding pump
using the Start/Stop key.
– P.C.B. LP633 recognizes Single-Needle pump although it is not
connected.
Applicable for SW 4.91/2.91 and higher if SN, ONLINE-HDF or
4008 HDF pump is connected (= ADKS active)
CPU1 fails to stop the corresponding blood pump.
CPU1 detects that the pump remains active.
– Control line (BPSST_VEN, X634L/B15) → X348V/1 not 0V as well
as (BPST_VEN, X634L/A15) → X348V/3 not 0V
– Acknowledgement (BPSB_VEN, X348V/6) → X633L/A13 not 0V
– IC16 on P.C.B. LP633 defective
– P.C.B. LP633 recognizes Single-Needle pump although it is not
connected.
Applicable for SW 4.91/2.91 and higher if SN, ONLINE-HDF or
4008 HDF pump is connected (= ADKS active)
Although the recognition of the venous blood pump (ADKS) is not
acknowledged, the 24-V supply voltage of the pump can be switched
off.
– Acknowledgement line (ADKS, X348V/7) → X633L/A10 not 12V
– Acknowledgement (BPSB_VEN, X348V/6) → X633L/A13 not 12V
– Acknowledgement (BPSB_VEN, X348V/6) → X632/B11 not 12V
– Online-HDF has already been switched on during the T1 test.
– IC16 on P.C.B. LP633 defective.