Fresenius Medical Care 4008 E Technical Manual page 179

Hemodialysis machine
Table of Contents

Advertisement

Part 8: 300 ml/min flow
CAL. FLOW 300 ml/min
Confirm
flow (300) = XXX
Change the digits in the "UF Rate"
window by pressing the
keys, until the actual value agrees
with the specified value (300).
Override
(Tone Mute)
DATA STORED
approx. 3 sec
Important:
If it is impossible to adjust the 300 – 500 – 800 flow volumes, or if problems caused by flow alarms
occur after the "calibrate flow" message has appeared, this can be caused by the setting of the
current rise pulse.
Note:
For the flow selected first, the message "DIASAFE-filling act." is displayed for 17 balancing
chamber switchings.
4-14
Fresenius Medical Care 4008 4/08.03 (TM)