Part 5.1: Calibrate arterial blood pump
CALIB. ART. BP-RATE
Confirm
key
calib. Art. BP-RATE
art. BP-Rate CHECK
back to menu ?
Note: Set the line diameter to 8 mm before starting the calibration procedure and press Start/Stop on the blood pump.
* The BP rate of 550 ml/min represents a default value. It can be changed using the
Part 5.2: Calibrate single needle blood pump rate
CALIB. SN. BP–RATE
Confirm
key
calib. sn. BP–RATE
SN. BP-Rate CHECK
back to menu ?
Note: Set the line diameter to 8 mm before starting the calibration procedure and press Start/Stop on the blood pump.
* The BP rate of 550 ml/min represents a default value. It can be changed using the
SN pump: lower switching point fixed to 75 mmHg
Stroke volume (ml)
Upper switching point
±
(mmHg)
7 mmHg
Confirm
key
Confirm
(B)P–Rate=550 ml/min
key
Set various delivery rates
on the arterial blood pump.
Confirm
The values on the display
key
and the BP must be
identical.
Select
(Esc)
key
Using a syringe, set the
pressure at the P-SN connector
according to the set
stroke volume (see table).
Confirm
(B)P–Rate=550 ml/min
key
Set various delivery rates
on the SN blood pump.
Confirm
The values on the display
key
and the BP must be identical.
Select
(Esc)
key
10
15
110
130
Fresenius Medical Care 4008 4/08.03 (TM)
(B)P–Rate=550 ml/min
Adjust a rate of
≥550 ml/min* on the
art. blood pump.
Override
(Tone Mute)
key
DATA STORED
approx. 3 sec
(+/–) keys.
Confirm
(B)P–Rate=550 ml/min
key
Set a rate of
≥550 ml/min*
on the SN blood pump.
Override
(Tone Mute)
key
DATA STORED
approx. 3 sec
(+/–) keys.
20
25
30
35
150
172
195
219
Select
(Esc)
key
Select
(Esc)
key
40
45
50
244
270
299
4-9