Lp 632 (Cpu 2) Dip Switch Array - Fresenius Medical Care 4008 E Technical Manual

Hemodialysis machine
Table of Contents

Advertisement

3.1.3
P.C.B. LP 632 (CPU 2) DIP switch array 1
LP632
SW1
ON
OFF
Switch
Switch 1
Switch 2
Switch 3
Switch 4
Switch 5
Switch 6
Switch 7
Switch 8
The basic position upon delivery is shown in italics.
For "not used" the switch must be set to OFF.
Note
DIP switches 3 and 8 permit to skip test steps which are requested by the
machine.
If the switches are set to the "can be skipped" position, it is important to know that
the operator can then bypass the automatic test of the safety systems.
The person demanding this switch position shall be solely responsible for such a
procedure.
1
2
3
4
5
6
7
8
Position
Function
ON
Not used
OFF
ON
T1 test, serial sequence
OFF
T1 test, parallel sequence
ON
Skip T1 test
OFF
T1 test mandatory
ON
Test service "ON" (hemodialysis not possible)
OFF
Test service "OFF" (automatic T1 test)
ON
Cyclic PHT every 2 minutes and indication of the test result (service)
OFF
Cyclic PHT every 12.5 minutes,
Alarm emission only with cyclical PHT alarm
ON
Cyclic PHT "ON"
OFF
Cyclic PHT "OFF"
ON
Air detector with PCB LP 450 without AD28
OFF
Air detector with PCB LP 450-2 or with AD28
ON
HDF test can be skipped
OFF
HDF test mandatory
SW2
Fresenius Medical Care 4008 4/08.03 (TM)
3-13