Fresenius Medical Care 4008 E Technical Manual page 281

Hemodialysis machine
Table of Contents

Advertisement

Part 10.1: Activate OCM measurement
OCM MEASUREMENT
Confirm
key
OCM Measurement: OFF
(default value:
+/–
OFF)
OCM Measurement: ON
+/–
Part 10.2: Activate OCM zero measurement
OCM ZERO MEASUREMENT
Confirm
key
OCM Zero Measure:ON
(default value:
+/–
ON)
OCM Zero Measure:OFF
+/–
Part 10.3: Set the OCM measurement delay time
OCM MEASURE DEL.TIME
Confirm
key
Delaytime: XX s
Press +/– to set the value indi-
cated in the table on the right
(1 – 70 s)
Tone Mute
key
DATA STORED
After approx. 3 sec
6-50
Fresenius Medical Care 4008 4/08.03 (TM)
Tone Mute
DATA STORED
key
Esc
key
Tone Mute
DATA STORED
key
Esc
key
Note
If the OCM zero measurement is set to "OFF", the OCM option is deactivated.
If the OCM option is reactivated ("ON"), an OCM pulse calibration must be
performed.
Tone Mute
DATA STORED
key
Esc
key
Tone Mute
DATA STORED
key
Esc
key
Esc
key
After
approx. 3 sec
After
approx. 3 sec
After
approx. 3 sec
After
approx. 3 sec
System
4008 H with advanced
4008 H
hydraulics
4008 H with ONLINEplus™
4008 S with advanced
4008 S
hydraulics
4008 S with ONLINEplus™
Time
15 Sec
65 Sec
18 Sec
67 Sec