Fresenius Medical Care 4008 E Technical Manual page 251

Hemodialysis machine
Table of Contents

Advertisement

Part 4: Set HDF dilution – only for ON-LINE-HDF (option)
HDF-DILUTION
Confirm
key
HDF-PRE-dilution
HDF-POST-dilution
Part 5: Set conductivity limit
SET CONDUCT. LIMIT
Confirm
key
Cd Limit: 12.8 mS/cm
Set the desired value
(12.8 to 14.0 mS/cm)
by pressing the
Override
key
DATA STORED
After approx. 3 sec
Part 6: Infosound cleaning program
INFO SOUND (C-PGM)
Confirm
key
Info-Sound: ON
Info-Sound: OFF
6-20
Fresenius Medical Care 4008 4/08.03 (TM)
Override
DATA STORED
key
Select
key
Override
DATA STORED
key
Select
key
Select
key
keys.
Override
DATA STORED
key
Select
key
Override
DATA STORED
key
Select
key
After
approx. 3 sec
After
approx. 3 sec
After
approx. 3 sec
After
approx. 3 sec