Fresenius Medical Care 4008 E Technical Manual page 116

Hemodialysis machine
Table of Contents

Advertisement

TSC MA
No.
Description
7.6
ONLINE plus ™ (and DIASAFE
MA
7.6.1
Filter life of DIASAFE
and ONLINE plus ™ filter
MA
7.6.2
Hydrophobic filters 111 and
184
MA
7.6.3
Substituate port (195) and
rinse port (194)
TSC
7.6.4
Line pinch valve 193 (ONL1)
TSC
7.6.5
Valve 39
Note:
This TSC item will not be
applicable if DIP switch
array 2, P.C.B. LP 632,
switch 5 is set to OFF.
7.7
OCM
No further technical safety checks and maintenance procedures must be per-
formed.
7.8
BPM 4008
Perform the technical measurement checks and maintenance procedures every 2
years (see section 2.2).
7.9
BTM 4008
No further technical safety checks and maintenance procedures must be per-
formed.
7.10
BVM 4008
No further technical safety checks and maintenance procedures must be per-
formed.
Expected value / function
®
plus )
®
plus
Check the filter life.
Filter life of DIASAFE
filter: 12 weeks or 100 treatments
Replace the filters.
Replace the O-rings.
Replace the line.
Check for proper function.
Fresenius Medical Care 4008 4/09.03 (TM)
®