Fresenius Medical Care 4008 E Technical Manual page 103

Hemodialysis machine
Table of Contents

Advertisement

Disinfection
PGM 1: –F–D–M–
T/°C
37
33
Rinsing clean
4 to 10 min
(Setup)
Start
(Disinfection key)
*Prep.: preparation phase:
Heater off
Set the level of the float switch chamber below the lower switching point
of the float switch by
1 balancing chamber changeover and
4 UF pump strokes.
Aspiration of disinfectant for 50 UF-pump strokes.
PGM 2: –F–HDIS–M–
T/°C
80
37
34,5
Rinsing clean
4 to 10 min
(Setup)
Start
(Disinfection key)
*Prep.: preparation phase:
Heater off
Set the level of the float switch chamber below the lower switching point
of the float switch by
1 balancing chamber changeover and
4 UF pump strokes.
Aspiration of disinfectant for 50 UF-pump strokes.
1-96
Fresenius Medical Care 4008 4/08.03 (TM)
Prep.*
Disinfection
approx.
10 to 20 min (Setup)
1 min
Mandatory rinse requested
Prep.*
Hot disinfection
Heating
approx.
approx.
10 to 20 min (Setup)
1 min
4 min
approx. 6 min
Mandatory rinse requested
Rinse chamber
Mandatory rinse
evacuation
10 strokes each
15 to 30 min (Setup)
approx. 1 min
Rinse chamber
Mandatory rinse
evacuation
10 strokes each
15 to 30 min (Setup)
approx. 1 min
t/min
End
t/min
End