Page 3
4008 HDF Technical Manual The Technical Manual contains all information necessary for performing maintenance and repair work. The 4008 HDF option reflects the latest state of technology and complies with the requirements of EN 60601-1 Assembly, extension, adjustment, modification or repair may only be carried out by the manufacturer or persons authorized by him.
Page 4
Fresenius Medical Care 4008 HDF 3/11.97 (TM)
The identification on the P.C.B. permits the user/technician to verify if the circuit diagram / component layout matches the P.C.B. actually installed. In general, this manual is subject to modification. Fresenius Medical Care 4008 HDF 3/11.97 (TM)
Page 6
ESD precautions (e.g. EN 100 015-1). During repair/troubleshooting in the hydraulic unit, protect the components from dialysate. Technical data: The technical data for the 4008 HDF option is to be found in Chapter 1 of the Operating Instructions. Fresenius Medical Care 4008 HDF 3/11.97 (TM)
LP 763 SSE serial interface extender ................4-39 Spare parts ........................How to use the spare parts catalog ................P.C.B.s ..........................Scales ..........................Substituate lift ......................... UF2 pump / blood pump (HDF) / valve V 126 ..............5-10 Fresenius Medical Care 4008 HDF 3/11.97 (TM)
Page 8
Fresenius Medical Care 4008 HDF 3/11.97 (TM)
Fig.: Flow diagram ......................Description of extended T1 test / error messages in T1 test ........1.2.1 Test UF-Function ......................Error messages during treatment and the HDF test ..........1-10 Error messages, substituate pump ................1-13 Fresenius Medical Care 4008 HDF 3/11.97 (TM)
Page 10
Fresenius Medical Care 4008 HDF 3/11.97 (TM)
– The substituate pump – The UF2 pump (electrically and hydraulically) (If the 4008 HDF test is not carried out then the connections are to be tested but once). The UF2 pump is tested cyclically during treatment. The electric control system of the UF2 pump is checked continuously.
Page 12
Fig.: Block diagram Operating system Safety system Scales Serial communication CPU 1 CPU 2 Serial communication Substituate UF2 pump pump Fresenius Medical Care 4008 HDF 3/11.97 (TM)
Page 13
Fig.: Flow diagram Fresenius Medical Care 4008 HDF 3/11.97 (TM)
Page 14
Fig.: Flow diagram Fresenius Medical Care 4008 HDF 3/11.97 (TM)
CPU 1 starts up the UF2 pump at a defined rate. CPU 2 controls the hydraulic and electrical function of the UF2 pump. CPU 2 blocks the control line of the UF2 pump and checks for standstill. UF2 counter check. Figure: Fresenius Medical Care 4008 HDF 3/11.97 (TM)
– Control line (UF_P2_CRTL, X632/B24) → X634R/C11 no HIGH pulse. F20 UF-Function Difference in pressure between UF pump stroke and UF2 pump stroke more than 20 %. – Stroke rate for UF pump or UF2 pump not set correctly. Fresenius Medical Care 4008 HDF 3/11.97 (TM)
Page 18
Check scales (for drifting). than –1 g, the respective UF pump test will be repeated. The UF pump test was repeat- ed more than 5 times. 1-10 Fresenius Medical Care 4008 HDF 3/11.97 (TM)
Page 19
(communication problems) of the UF pump is greater than 10 ml. F334 UF-failure Interval between two strokes Check CPU 1 (defect). of the UF2 pump less than 220ms. 1-11 Fresenius Medical Care 4008 HDF 3/11.97 (TM)
Page 20
2g. F349 HDF-failure The delivery rate correction Check substituate bag factor could still not be deter- (weight fluctuations) mined after 20 minutes treat- Check HDF control system ment time. (greater fluctuations) 1-12 Fresenius Medical Care 4008 HDF 3/11.97 (TM)
Page 21
Not defined E.10 Not defined E.11 Not defined E.12 Failure, speed monitoring (hall sensor) E.13 Failure, monitoring system, current sensing resistors E.14 Failure, monitoring system, current sensing resistors E.15 Failure, speed monitoring system 1-13 Fresenius Medical Care 4008 HDF 3/11.97 (TM)
Page 22
1-14 Fresenius Medical Care 4008 HDF 3/11.97 (TM)
The technical safety inspections stipulated for the hemodialysis machine must be carried out in addition to these technical saftey checks. The technical safety checks are to be recorded in the equipment log. A technical safety checks report is to be found at the page 2-3. Fresenius Medical Care 4008 HDF 3/11.97 (TM)
Page 24
Fresenius Medical Care 4008 HDF 3/11.97 (TM)
The correct completion of the listed work and the correctness of statements made are hereby certified. Signature, technician: Signature, customer: ......, date ............. , date ..................................Fresenius Medical Care 4008 HDF 3/11.97 (TM)
Page 26
Fresenius Medical Care 4008 HDF 3/11.97 (TM)
Enter the number of strokes by pressing Confirm Select pulse-amount = 60 keys. Press the Override key. Single Needle –10 –20 Override –30 Test –40 Prime Text display for approx 3 sec. Confirm Select ACKNOWLEDGED Fresenius Medical Care 4008 HDF 3/11.97 (TM)
Page 32
Text display after completion. Confirm Select UF-Pump 2 Check delivered quantity. If necessary adjust UF 2 pump and re- peat process. Delivery rate: 1 ml/stroke (60 strokes = 60 ml) Tolerance: ± 1 %. Fresenius Medical Care 4008 HDF 3/11.97 (TM)
The scales must be travelled out to such an extent that neither of the two end switches are actuated (middle position). The two leakage currents are to be measured each time with different mains connection poling. The two leakage currents are added together and give the summarized leakage current. Fresenius Medical Care 4008 HDF 3/11.97 (TM)
Switch ON? Open housing of scales at the back. Fresenius Medical Care Set calibration switch for the scales in ON position. Press the Override key. Single Needle –10 –20 Override –30 Test –40 Prime Fresenius Medical Care 4008 HDF 3/11.97 (TM)
Page 35
No weight is to be on the scales. Press the Override key. Single –10 Needle –20 Override –30 Test –40 Prime Text display. Confirm Select HDF-DISPL shows 0 ? Scales display. Fresenius Medical Care Fresenius Medical Care 4008 HDF 3/11.97 (TM)
Page 36
Fresenius Medical Care scales. Press the Override key. Single Needle –10 –20 Override –30 Test –40 Prime Text display. Select Confirm DISPL shows UNIT “g”? Scales display: 5000 g. Fresenius Medical Care 3-10 Fresenius Medical Care 4008 HDF 3/11.97 (TM)
Page 37
HDF-SCALE unloaded ? Remove calibration weight. Press the Override key. Single Needle –10 –20 Override –30 Test –40 Prime Brief text display. Confirm Select SCALE CALIBRAT. done Confirm Select CALIB. HDF-SCALE Text display. 3-11 Fresenius Medical Care 4008 HDF 3/11.97 (TM)
Page 38
Set service switch to OFF (bottom). Switch on hemodialysis machine. Hang calibration weight PT6 – 5000 g on scales. Scales display: 5000 g Fresenius Medical Care Remove calibration weight. Calibration of 4008 HDF scales completed. 3-12 Fresenius Medical Care 4008 HDF 3/11.97 (TM)
Needle –10 –100 UF Ziel –20 Überbrücken –200 –30 Test –300 –40 Vorbereiten UF Restzeit h:min Dialyse Start Alarm Prog. Ton Aus Rate: ml/h Rate: ml/min (Bolus: ml) (Ø: mm) :h.min) Bolus 3-13 Fresenius Medical Care 4008 HDF 3/11.97 (TM)
Page 40
IC17 X348V X348V_SN XHDF Turn back the potentiometer P1 on the LP 754 by half this number. The LED D26 will illuminate again. IC14 IC17 X348V X348V_SN XHDF 3-14 Fresenius Medical Care 4008 HDF 3/11.97 (TM)
Vol : l Start Stop Display 250 Rate: ml/min (Ø: mm) Apply a pressure of 250 mmHg to the SN pressure transducer. Reset Press the Start/Stop key. Vol : l Start Stop 3-15 Fresenius Medical Care 4008 HDF 3/11.97 (TM)
Balancing out has been completed. X348V_SN XHDF 3.5.2 Checking SN switchover points Bottom switchover point: fixed at 75 mmHg (± 6 mmHg) Top switchover points (± 6 mmHg): Stroke volumes (ml) Top switchover point 3-16 Fresenius Medical Care 4008 HDF 3/11.97 (TM)
Repair instructions Important: If the substituate lift has to be dis- mounted for repair the seal is to be checked and if necessary renewed before fitting the lift again. Drip water protection! 3-17 Fresenius Medical Care 4008 HDF 3/11.97 (TM)
Page 44
3-18 Fresenius Medical Care 4008 HDF 3/11.97 (TM)
Page 46
Fresenius Medical Care 4008 HDF 3/11.97 (TM)
Page 47
Vol : l Start Stop XWAAGE X348v XHDF Substituate pump X348v LP 747 distribution board X637B X637B XHDF X637C X637C XA1P LP 763 SEE serial interface XA4V0 extender UF2 pump V126 LP 630 motherboard Fresenius Medical Care 4008 HDF 3/11.97 (TM)
Page 48
Fresenius Medical Care 4008 HDF 3/11.97 (TM)
Page 49
W1IW2 X63Z X637B X637C LP 763 LP 747 ON/OFF SEE serial distribution no assignment interface extender board XA1P XA4V0 Switch substituate lift UF2 pump V 126 Fresenius Medical Care 4008 HDF 3/11.97 (TM)
Page 50
Fresenius Medical Care 4008 HDF 3/11.97 (TM)
Page 51
Fig.: Block circuit diagram, mains voltage 4008 HDF Main Standby- trans- trans- former former T 2,5 A T 2,5 A Mains relay Motor relay T 2,5 A 1 µF Mains Mains Mains 511 K T 2,5 A Auxiliary plug filter...
Page 52
Fresenius Medical Care 4008 HDF 3/11.97 (TM)
The keyboard comprises 3 single non-illuminated keys and 1 key with integrated LED (HDF). The keys are soldered directly to the P.C.B.. The user receives a acknowledgement through a spring contact when the keys are pressed. Fresenius Medical Care 4008 HDF 3/11.97 (TM)
Page 54
4-10 Fresenius Medical Care 4008 HDF 3/11.97 (TM)
– X192, connection to pressure transducer – X348, connection to dialysis monitor Hexswitches on the LP 754: No function Single needle blood pump / 4008 HDF / ON LINE HDF 2 – A No function Blood pump stop alarm (15, 30 secs.) No function...
Page 60
AN 7 and calibrated for zero and steepness by the software. Afterwards the measuring signal is modulated to pulse width through port P1.3 and transmitted through a subsequent DA converter IC 11 (5/6/7) to the monitor. 4-16 Fresenius Medical Care 4008 HDF 3/11.97 (TM)
R 20 serves as a pull-up resistor. A defined output level is obtained with the IC 5/14 in high ohmic state, caused by pressing the Stop key (high level on IC 5/15). 4-23 Fresenius Medical Care 4008 HDF 3/11.97 (TM)
Page 68
R 11, R 12 and the B-E path from T 4. T 4 blocks after approx. 500 ms. The output transistor must be linked by IC 4 before this time has elapsed otherwise both flip-flops will be reset again. 4-24 Fresenius Medical Care 4008 HDF 3/11.97 (TM)
LP 761 HDF drive keyboard 4.4.1 Description ● General The keys for controlling the substituate lift are found on this P.C.B.. 4-29 Fresenius Medical Care 4008 HDF 3/11.97 (TM)
Page 74
4-30 Fresenius Medical Care 4008 HDF 3/11.97 (TM)
Page 75
1 X1 SUB_A 3 X1 SUB_C 5 X1 6 X1 1994 NAME HDF-ANTRIEB-TASTATUR BLATT 04.08. OTT 1(1) 05.08. ASCH HDF-MOTOR-KEYBOARD LP761 673 766 MEDIZIN- TECHN. X1 geandert SYSTEME ERSATZ ERSATZ FORMAT AENDERUNG 4-31 Fresenius Medical Care 4008 HDF 3/11.97 (TM)
Page 76
4-32 Fresenius Medical Care 4008 HDF 3/11.97 (TM)
Page 77
LP 761 Component layout diagram 4-33 Fresenius Medical Care 4008 HDF 3/11.97 (TM)
Page 78
4-34 Fresenius Medical Care 4008 HDF 3/11.97 (TM)
IC 1 from 5 V to RS 232 level. The signal arrives at the dialysis machine through plug X1 pin 5 (TXD). SP1 (ML78L05) converts the 12 V into 5 V which serves as the supply voltage for LP 762. 4-35 Fresenius Medical Care 4008 HDF 3/11.97 (TM)
Page 80
4-36 Fresenius Medical Care 4008 HDF 3/11.97 (TM)
LP 752 takes over the connection of the 4008 monitor signals to the COMMCO P.C.B. LP 729. In order to be able to also continue the linking to COMMCO/FINESSE in 4008 HDF machines the circuit section of the LP 752 P.C.B. is taken over in the LP 763 P.C.B..
Page 86
4-42 Fresenius Medical Care 4008 HDF 3/11.97 (TM)
Table of contents 5 Spare parts Section Page How to use the spare parts catalog ................P.C.B.s ........................... Scales ..........................Substituate lift ....................... UF2 pump / blood pump (HDF) / valve V126 ............. 5-10 Fresenius Medical Care 4008 HDF 3/11.97 (TM)
Page 94
Fresenius Medical Care 4008 HDF 3/11.97 (TM)
The equipment code must be entered in the machine record. All modifications must be entered in the machine record. Update the equipment code, if necessary, to ensure that the correct spare parts will be ordered. Fresenius Medical Care 4008 HDF 3/11.97 (TM)