Fresenius Medical Care 4008 E Technical Manual page 282

Hemodialysis machine
Table of Contents

Advertisement

Part 10.4: Set OCM correction factor
Note
Do not change the default value.
Note
The 4008 H/S offers the theoretical possibility of defining a correction value for
all clearance measurements. The intention is to be able to adapt the system to
to the latest state of technology resulting from the on-going scientific discussi-
on. According to the current standard of knowledge, this correction value has
to be set to "1" (factory setting) for Fresenius polysulfone membranes. The in-
dicated accuracy specifications are valid only for Fresenius polysulfone mem-
brane combined with this correction value setting of "1".
OCM CLR CORR FACTOR
Confirm
key
ClrCorr Factor:1.000
Display value
is
default value
Tone Mute
key
DATA STORED
After approx. 3 sec
Part 10.5: Set OCM baseline difference
Note
Do not change the default value.
OCM BASELINE DIFF.
Confirm
key
Baseline Diff:0.015
Display value
is
default value
Tone Mute
key
DATA STORED
After approx. 3 sec
Esc
key
Esc
key
Fresenius Medical Care 4008 4/08.03 (TM)
6-51