Fresenius Medical Care 4008 E Technical Manual page 43

Hemodialysis machine
Table of Contents

Advertisement

F05 pos. Pressure
F06 pos. Pressure
F07 pos. Pressure
F08 pos. Pressure
F09 pos. Pressure
F10 pos. Pressure
F11 pos. Pressure
1-36
Fresenius Medical Care 4008 4/08.03 (TM)
Test detuning results in a change in the measuring range of more than
95 mmHg (60 sec).
– The operational amplifier (IC2) on P.C.B. LP 632 is defective.
– Acknowledgement (P_DIAL, X633L/B6) → X632/A29, change in
voltage too large.
– Detuning defective (P_DETADJ, X632/C20) → X633R/C22.
– The balancing chamber valve V36 or V38 (old water valve) is
leaking.
Test detuning results in a change in the measuring range of less than
85 mmHg (60 sec).
– The D-A converter (IC11) on P.C.B. LP 632 is defective.
– Acknowledgement (P_DIAL, X633L/B6) → X632/A29, change in
voltage insufficient.
– Detuning defective (P_DETADJ, X632/C20) → X633R/C22.
– V26 is leaking.
After detuning in the test there is a difference (P.diff > ±9 mmHg)
between the display and the differential amplifier.
– The voltage divider R23/R9 or the operational amplifier IC2 is
defective.
– The operational amplifier IC1/IC3 is defective.
– The balancing chamber valve V36 or V38 (old water valve) is
leaking.
Test detuning results in a change in the measuring range of more than
400 mmHg (20 sec).
– The operational amplifier (IC2) on P.C.B. LP 632 is defective.
– Acknowledgement (P_DIAL, X633L/B6) → X632/A29, change in
voltage too large.
– Detuning defective (P_DETADJ, X632/C20) → X633R/C22.
Test detuning results in a change in the measuring range of less than
350 mmHg (20 sec).
– The D-A converter (IC11) on P.C.B. LP 632 is defective.
– Acknowledgement (P_DIAL, X633L/B6) → X632/A29, change in
voltage insufficient.
– Detuning defective (DIAL_DET_ADJ, X632/C20) → X633R/C22.
The second working point (116 digits) of the operational amplifier
cannot be set correctly.
– The D-A converter (IC11) on P.C.B. LP 632 is defective.
– The operational amplifier (IC1/IC3) on P.C.B. LP 632 is defective.
Change in the dialysate pressure after closing of the solenoid valve V43
(zero point change of ±20 mmHg within 15 sec).
– The solenoid valve V24B is not closed.
– The balancing chamber valve V36 or V38 (old water valve) is
leaking.
The balancing chamber is switched to passage during this test sequence. V43, V24B
and V26 are closed; V24 is open.