Water alarm
Water alarm
Flow alarm
Upper Flow Alarm
UF-Pump failed
UF-Pump 2 failed
Dial. Valve failed
Bypass Valve failed
During a cleaning program
in RI I to II, HR I to III, Dis I to V.
The float switch transmits the "no water available" message for more
than 10 seconds.
The balancing chamber has stopped; V41 is permanently open.
Upon error elimination, the message is automatically cleared.
During a cleaning program
in RI I to II, HR I to III, Dis I to V.
For more than 30 seconds, the float switch fails to signal that water is
required (not applicable to recirculation programs).
The machine continues to run.
Upon error elimination, the message is automatically cleared.
During a cleaning program
in RI I to II, HR I to III, Dis I to V.
A current rise pulse is not recognized for more than 12 seconds.
The machine continues to run at "Eigentakt" (10 seconds).
Upon error elimination, the message is automatically cleared.
During a cleaning program
in RI I to II, HR I to III, Dis I to V.
The cleaning flow increases to > 1000 ml/min. The program has
stopped.
The error can be acknowledged by pressing the respective cleaning
program key.
During a cleaning program
in RI I to II, HR I to III, Dis I to V.
The UF pump has stopped or the rate deviates (2800 ml/h < UFR <
6000 ml/h). The program has stopped.
The error can be acknowledged by pressing the respective cleaning
program key.
During a cleaning program
in RI I to II, HR I to III, Dis I to VI.
The UF2 pump has stopped (applicable only to machines with 4008
HDF).
The error can be acknowledged by pressing the respective cleaning
program key.
During a cleaning program
in RI I to II, HR I to III, Dis I to V.
V24 or V24B is closed although it should be open.
The program has stopped.
The error message can be acknowledged by pressing the respective
program key.
During a cleaning program
in RI I to II, HR I to III, Dis I to V.
V26 is closed although it should be open.
The program has stopped.
The error message can be acknowledged by pressing the respective
program key.