Part 18: Initialize NOVRAM, reset mandatory rinse, reset V84 malfunction
NOVRAM
Confirm
key
Init NOVRAM
Reset MAND. RINSE
Reset CDS-MAND RINSE
Reset V84
back to menu ?
Confirm
key
Confirm
key
audible alarm
Confirm
Are you sure ?
key
Override
(Tone Mute)
key
ACKNOWLEDGED
approx. 3 sec
BACK TO MAIN MENU ?
audible alarm
Confirm
Are you sure ?
key
Override
(Tone Mute)
Confirm
key
key
ACKNOWLEDGED
approx. 3 sec
BACK TO MAIN MENU ?
Fresenius Medical Care 4008 4/08.03 (TM)
audible alarm
Are you sure ?
Override
(Tone Mute)
key
ACKNOWLEDGED
approx. 3 sec
BACK TO MAIN MENU ?
audible alarm
Are you sure ?
Override
(Tone Mute)
Select
key
(Esc)
key
ACKNOWLEDGED
approx. 3 sec
BACK TO MAIN MENU ?
Select
(Esc)
key
Note:
The "NOVRAM" menu option can be entered
only if the DIP-Switch1/DIP-Sw.Array2/LP631 is
activated in the calibration mode. To this end
and depending on its initial position, the switch
must be actuated once and then be reset to
its initial position.
Select
(Esc)
key
Select
(Esc)
key
4-33