Fresenius Medical Care 4008 E Technical Manual page 105

Hemodialysis machine
Table of Contents

Advertisement

PGM 4: –F–HDIS–M–HR–
T/°C
80
37
Rinsing clean
4 to 10 min
(Setup)
Start
(Disinfection key)
*Prep.: preparation phase:
Heater off
Set the level of the float switch chamber below the lower switching point
of the float switch by
1 balancing chamber changeover and
4 UF pump strokes.
Aspiration of disinfectant for 50 UF-pump strokes.
PGM 5: –F–D(F)–M–
T/°C
37
33
Rinsing clean
4 to 10 min
(Setup)
Start
(Disinfection key)
*Prep.: preparation phase:
Heater off
Set the level of the float switch chamber below the lower switching point
of the float switch by
23 UF pump strokes.
Aspiration of disinfectant for
32 concentrate pump strokes à 330 steps.
1-98
Fresenius Medical Care 4008 4/08.03 (TM)
Hot
Mandatory rinse
Prep.*
disinfection
Heating
10 to 20 min
15 to 30 min
approx.
approx.
(Setup)
(Setup)
1 min
4 min
approx. 6 min
Mandatory rinse requested
Prep.*
Disinfection
approx.
10 to 20 min (Setup)
1 min
Mandatory rinse requested
Rinse chamber
Hot
evacuation
rinsing
Heating
10 strokes each
15 to 30 min
approx.
approx. 1 min
(Setup)
4 min
approx. 6 min
Rinse chamber
Mandatory rinse
evacuation
10 strokes each
15 to 30 min (Setup)
approx. 1 min
t/min
End
t/min
End