Fresenius Medical Care 4008 E Technical Manual page 114

Hemodialysis machine
Table of Contents

Advertisement

TSC MA
No.
Description
7
Options
7.1
bibag
MA
7.1.1
bibag
MA
7.1.2
PSW 134
7.2
DIASAFE
MA
7.2.1
DIASAFE filter life
MA
7.2.2
Hydrophobic filter 111
MA
7.2.3
O-rings in the dialysate
couplings of the DIASAFE
7.3
DIASAFE
MA
7.3.1
DIASAFE
MA
7.3.2
Hydrophobic filter 111
7.4
4008 HDF
TSC
7.4.1
Check the delivery rate of
the 2
®
®
connector
®
plus
®
plus filter life
nd
UF pump.
Fresenius Medical Care 4008 4/09.03 (TM)
Expected value / function
Replace the O-rings.
Check the switching pressure.
The maximum switching pressure is
100 mbar + 10 mbar.
Check the filter life.
Filter life: 12 weeks
Replace the filter.
Replace the O-rings.
Check the filter life.
Filter life: 12 weeks
Replace the filter.
Collect 60 ml of dialysate in the dialysis mode
using an appropriate measuring cylinder.
60 strokes = 60 ml (±0.5 ml)
If necessary, correct the value.
2-9