Error description:
Error message
F01 Temperature
F02 Temperature
F03 Temperature
F04 Temperature
F05 Temperature
F06 Temperature
F07 Temperature
F08 Temperature
F09 Temperature
F10 Temperature
F95 Temperature
Description
The temperature measuring range is not set to hemodialysis.
– Control line (HOTRINSE, X634R/C24) → X639/A20, 0 V are miss-
ing.
– Acknowledgement (HOTRINSE, X634R/C24) → X632/A26, 0 V are
missing.
The actual temperature is less than 35.0 °C (test running time > 15
minutes).
– Calibrate the temperature.
– The heater rod failed.
– Acknowledgement (T_DIAL1, X633L/B16) → X632/A24, voltage got
stuck.
The actual temperature is higher than 39.0 °C (test running time > 15
minutes).
– Calibrate the temperature.
– The regulating sensor (NTC-2) is defective.
– Acknowledgement (T_DIAL1, X633L/B16) → X632/A24, voltage got
stuck.
The temperature failed to stabilize within 15 minutes.
– Acknowledgement (T_DIAL1, X633L/B16) → X632/A24 is steadily
changing (change > 0.3 °C/15 sec).
Detuning in positive direction not higher than 3 °C (10 sec).
– Acknowledgement (T_DIAL1, X633L/B16) → X632/A24, change in
voltage insufficient.
– Detuning (T_DETADJ, X632/A23) → X633R/C21 insufficient.
The monitor sensor indicates a constant value.
– NTC-3 defective.
The test release is missing (max. test running time is 10 minutes).
– Run-time problem (software).
CPU1 failed to transmit a Bibag status message within 3 sec.
– Run-time problem (software).
Bibag NTC_BIB detuning not higher than 1 °C.
– Acknowledgement (NTC_BIB, X633R/C15) → ADW on P.C.B.
LP 633, change in voltage insufficient.
– Detuning (BIBAG_TE, X634R/A13) → X633R/A20 insufficient.
Bibag temperature display outside of measuring range (15 to 45 °C).
– Acknowledgement (NTC_BIB, X633R/C15) → ADW on P.C.B.
LP 633.
System error.
Fresenius Medical Care 4008 4/08.03 (TM)