Fresenius Medical Care 4008 E Technical Manual page 121

Hemodialysis machine
Table of Contents

Advertisement

4
Ultrafiltration system and membrane pumps
TSC 4.1
UF pump, 1 stroke = 1 ml, 60 strokes = 60 ml ±0.5 ml ........... Measured value:________
MA
4.2
Average balancing chamber volume 30 ± 1 ml....................... Measured value:________
MA
4.3
Concentrate pump calibration volume removal /
number of strokes ................................................................... Measured value:________
MA
4.4
Bicarbonate pump calibration volume removal /
number of strokes ................................................................... Measured value:________
5
Dialysis mode
MA
5.1
Expected temperature 37 °C ± 0.5 °C .................................... Measured value:________
MA
5.2
Temperature display 37 °C ± 0.5 °C ....................................... Measured value:________
MA
5.3
Dialysate flow check
800 ml/min (desired value: 765 to 837 ml/min) ................... Measured value:________
500 ml/min (desired value: 471 to 528 ml/min) ................... Measured value:________
300 ml/min (desired value: 279 to 321 ml/min) ................... Measured value:________
MA
5.4
Dialysate pressure
– Check zero point with flow off ........................................................................................
– Slope checked ................................................................................................................
TSC 5.5
Conductivity display checked with reference meter ...........................................................
If the bibag
– CD system ........................................................................... Measured value:________
– CD ref. .................................................................................. Measured value:________
6
Extracorporeal components
MA
6.1
Arterial pressure displayed checked with reference meter ................................................
MA
6.2
Venous pressure displayed checked with reference meter ...............................................
TSC 6.3
Blood pumps: blood pump rate checked (calibration program: BP-Rate CHECK) ............
TSC 6.4
SN switching pressure checked according to table in TM .................................................
TSC 6.5
Blood pump stop alarm checked .......................................................................................
TSC 6.6
Venous line clamp closes after blood alarm ......................................................................
TSC 6.7
Pressure of about 2 bar in the venous bubble catcher ......................................................
Pressure must not drop by more than 0.1 bar within 3 minutes. .......................................
7
Options
®
7.1
bibag
®
MA
7.1.1
bibag
connector, O-rings replaced ...................................................................................
MA
7.1.2
Switching pressure of PSW134 checked,
100 mbar, + 10 mbar ............................................................... Measured value:________
7.2
DIASAFE
MA
7.2.1
DIASAFE filter life checked ................................................................................................
MA
7.2.2
Hydrophobic filter 111 replaced .........................................................................................
MA
7.2.3
O-rings in the dialysate couplings of the DIASAFE replaced ............................................
7.3
DIASAFE
MA
7.3.1
DIASAFE
MA
7.3.2
Hydrophobic filter 111 replaced .........................................................................................
7.4
4008 HDF
nd
TSC 7.4.1
2
UF pump 1 stroke = 1 ml, 60 strokes = 60 ml ± 0.5 ml ..... Measured value:________
7.5
ON-LINE-HDF (and DIASAFE)
MA
7.5.1
Filter life of the DIASAFE and ON-LINE filter checked ......................................................
MA
7.5.2
Hydrophobic filter 111 replaced .........................................................................................
MA
7.5.3
O-rings in the dialysate couplings of the DIASAFE replaced ............................................
MA
7.5.4
HDF pump rotor checked (smooth running, wear) ............................................................
MA
7.5.5
Fastening strap for Luer-lock checked ..............................................................................
TSC 7.5.6
Substituate pump ................................................................................................................
(part no. 672 521 1) with DC motor:
volume delivered by the pump checked ............................... desired/actual:____/____
or
(part no. 674 982 1) with stepper motor:
pump rate checked (calibration program: HDF-P.-Rate CHECK)
2-16
Fresenius Medical Care 4008 4/09.03 (TM)
®
option is used, connect a bibag
®
plus
®
plus filter life checked .........................................................................................
®
!