Lp 631 (Cpu 1) Dip Switch Array - Fresenius Medical Care 4008 E Technical Manual

Hemodialysis machine
Table of Contents

Advertisement

3.1.2
P.C.B. LP 631 (CPU 1) DIP switch array 2
SH1
Switch / Position
SW 1
ON
OFF
SW 2
SW 3
ON
ON
OFF
ON
ON
OFF
OFF
OFF
SW 4
ON
OFF
SW 5
ON
OFF
SW 6
ON
OFF
SW 7
ON
OFF
SW 8
ON
OFF
The basic position upon delivery is shown in italics.
For "not used" the switch must be set to OFF.
3-12
Fresenius Medical Care 4008 4/08.03 (TM)
SH2
1
2
3
ON
OFF
Function
CAL mode
Mode 0
Mode 1
Ext. alarm input
Invalid
RO system
Patient bell
Ext. alarm
Remote control
System with remote control
System without remote control
COMMCO LP 763 or LP 758 or LP 729
Enabled
Disabled
COMMCO
Special record
Standard record
Hydraulics test (not CDS)
Active
Inactive
Central delivery system
Installed
Not installed
LP631
4
5
6
7
8