Fresenius Medical Care 4008 E Technical Manual page 104

Hemodialysis machine
Table of Contents

Advertisement

PGM 3: –F–D–M–HR
T/°C
80
37
33
Prep.*
Rinsing clean
4 to 10 min
approx.
10 to 20 min
(Setup)
1 min
Start
Mandatory rinse requested
(Disinfection key)
*Prep.: preparation phase:
Heater off
Set the level of the float switch chamber below the lower switching point
of the float switch by
1 balancing chamber changeover and
4 UF pump strokes.
Aspiration of disinfectant for 50 UF-pump strokes.
Dis-
Mandatory rinse
infection
Heating
15 to 30 min
approx.
(Setup)
(Setup)
4 min
approx. 6 min
Fresenius Medical Care 4008 4/08.03 (TM)
Rinse chamber
Hot rinsing
evacuation
10 strokes each
15 to 30 min
approx. 1 min
(Setup)
End
t/min
1-97

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

4008 b4008 h4008 s

Table of Contents