Part 3.1: Programmable mixing ratio (dilution)
SETUP DILUTION
Confirm
key
+/–
VARIABLE SETTING
6-44
Fresenius Medical Care 4008 4/08.03 (TM)
Caution
The operator or technician is informed about his duty of care to enter the com-
ponent parameters and settings correctly or to check them and to set the CD
alarm window to the expected conductivity value.
When using the programmable mixing ratio (dilution), make sure to use the
right concentrate.
®
Using a bibag
in combination with the programmable mixing ratio (dilution) is
not allowed.
Only enter authorized, programmable mixing ratios (dilutions).
Confirm
B-Comp[part]:0.800
key
Set the desired value
by pressing the +/– keys.
ACKNOWLEDGED
RO[part] : 30.000
Set the desired value
by pressing the +/– keys.
ACKNOWLEDGED
HCO3–[mmol]: 25
Set the desired value
by pressing the +/– keys.
ACKNOWLEDGED
Na(B)[mmol]: 25
Set the desired value
by pressing the +/– keys.
ACKNOWLEDGED
Esc
key
(0.800 – 2.500)
Tone Mute
key
After approx. 3 sec
Esc
key
(30.000 – 45.000)
Tone Mute
key
After approx. 3 sec
Esc
key
(25 – 45)
Tone Mute
key
After approx. 3 sec
Esc
key
(25 – 80)
Tone Mute
key
After approx. 3 sec