Rinse Failure F13
In case of F07, F08 and F13, the "DO NOT SWITCH OFF !!" message can, in addition, be
alternately displayed.
However, this message is displayed only if a mandatory rinse program is requested, since the
concentrate and bicarbonate lines still have to be emptied before the machine is switched off.
Rinse-free program with following Dis or HDIS or mandatory rinse as
individual program in Dis I to V.
Pressure drop during the monitoring phase on PSW_104 (S124) (bicar-
bonate) or PSW_102 (S123) (concentrate) or pressure build-up impos-
sible.
– Check switching point of pressure switch.
– Check loading pressure.
(Possibly splinter or contamination in orifice 151; remove tube and
blow through tube from both ends)
– Check negative pressure and orifice (89).
(For this purpose, remove and purge the tubing from both
ends)
– Check check valve (117/118) and filter (119/120).
– Check CDS valve (102/104).
– Verify tightness of CDS path.
– Check acknowledgement of pressure switch on P.C.B. LP 633:
Bicarbonate: X633L/ A19
Concentrate: X633L/ A20
– Cartridge filter upstream of degassing pump clogged or wrong filter
(filter for disinfectant container) installed. Filters can be distin-
guished by different adapters.
Fresenius Medical Care 4008 4/08.03 (TM)
1-57