Fresenius Medical Care 4008 E Technical Manual page 255

Hemodialysis machine
Table of Contents

Advertisement

Part 12: Activation of standard UF data
ACTIV. STRD UF-DATA
Confirm
key
strd UF-DATA: NO
strd UF-DATA: YES
Part 13: Set priming time
SET STD. PRIME-TIME
Confirm
key
Prime-Time = 2 min
Set the desired time
(1 to 5 min) by pressing
the
keys.
Override
key
DATA STORED
After approx. 3 sec
Part 14: Warning sound I/O key
SOUND I/O SWITCH
Confirm
key
I/O-Warnsound: ON
I/O-Warnsound: OFF
6-24
Fresenius Medical Care 4008 4/08.03 (TM)
Override
key
Select
key
Override
key
Select
key
UF-Rate:
750 ml/h
UF-Goal:
3000 ml
Time Left:
4 h
Select
key
Override
key
Select
key
Override
key
Select
key
DATA STORED
approx. 3 sec
DATA STORED
approx. 3 sec
DATA STORED
approx. 3 sec
DATA STORED
approx. 3 sec
After
After
After
After