Fresenius Medical Care 4008 E Technical Manual page 41

Hemodialysis machine
Table of Contents

Advertisement

Positive pressure holding test
Test description:
Valves V24, V24B and V26 are checked for proper function (mechanical).
Test of the TMP unit by detuning it electronically in positive direction.
With the dialysate flow turned off, positive pressure is applied to the balancing system. The actual
value of the dialysate pressure transducer is now monitored for a defined period of time.
Test of the pump segment of P97.