Fresenius Medical Care 4008 E Technical Manual page 196

Hemodialysis machine
Table of Contents

Advertisement

Part 16: Calibrate BIBAG values (optional)
CALIB. BIBAG VALUES
Confirm
key
BIBAG Temp.-Adjust
BIBAG Temp.-Check
BIBAG Cond.-Adjust
BIBAG Cond.-Check
back to menu ?
Confirm
key
Confirm
act temp = XX.X°C
key
Confirm
connect 104Ω±0.1%
key
Connect the calibration resistor
(104Ω) to X108.
Override
(Tone Mute)
key
ACKNOWLEDGED
approx. 3 sec
connect 56.2Ω±0.1%
Connect the calibration resistor
(56.2Ω) to X108.
Override
(Tone Mute)
key
DATA STORED
approx. 3 sec
Confirm
act.cond: XX.XmS/cm
key
Confirm
key
Fresenius Medical Care 4008 4/08.03 (TM)
connect 10kΩ±0.1%
Connect the calibration resistor
*
(10 kΩ) to X107.
Select
(Esc)
key
Select
(Esc)
key
DATA STORED
Select
(Esc)
key
**
Select
(Esc)
key
with test plug: 25 °C
*
**
without test plug: 45,7 mS/cm or 84,5 mS/cm
refer also to the Technical Manual for the biBag
Select
(Esc)
key
Override
(Tone Mute)
key
approx. 3 sec
4-31