Fresenius Medical Care 4008 E Technical Manual page 112

Hemodialysis machine
Table of Contents

Advertisement

TSC MA
No.
Description
4
Ultrafiltration system and membrane pumps
TSC
4.1
Check the delivery volume of
the UF pump.
MA
4.2
Check the balancing cham-
ber volume.
MA
4.3
Test the concentrate pump
volume or compare it with an
appropriate reference.
MA
4.4
Test the bicarbonate pump
volume or compare it with an
appropriate reference.
5
Dialysis mode
MA
5.1
Temperature
MA
5.2
Temperature display
(not applicable for 4008 B /
4008 S)
MA
5.3
Verify the dialysate flow
300/500/800 ml/min
MA
5.4
Dialysate pressure
TSC
5.5
Verify the conductivity dis-
play
Expected value / function
In the dialysis mode collect 60 ml of dialysate in
an appropriate graduated cylinder.
60 strokes = 60 ml (±0.5 ml).
Correct the setting of the UF pump, if necessary.
Collect the volume of two consecutive balancing
chamber switchings.
The average volume must be 30 ml ±1 ml.
Adjust according to calibration instructions, if ne-
cessary.
Adjust according to calibration instructions, if ne-
cessary.
Use a reference meter connected between the
dialyzer couplings to verify that the temperature is
37 °C ±0.5 °C.
Correct any deviations with the calibration pro-
gram.
The temperature shown on the monitor front pa-
nel must be 37 °C ±0.5 °C.
Correct any deviations with the calibration pro-
gram.
Collect fluid on the drain using a measuring cylin-
der.
800 ml/min (desired value: 765 to 837 ml/min)
500 ml/min (desired value: 471 to 528 ml/min)
300 ml/min (desired value: 279 to 321 ml/min)
Adjust according to calibration instructions, if ne-
cessary.
Perform a TMP test according to the calibration
instructions.
(part 14 CAL. DIAL. PRESSURE)
When the bibag
®
®
bibag
!
Measure the conductivity with a reference meter
connected bewteen the dialyzer couplings.
The conductivity measured must agree with the
value on the reference meter.
Correct any deviations with the calibration pro-
gram.
Fresenius Medical Care 4008 4/09.03 (TM)
option is used, connect a
2-7

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

4008 b4008 h4008 s