Calibration for NA, CAT, and VVM Modes
— Press Cal 5 then
— For 1-port measurements:
— For 2-port measurements:
With a Normalization Cal, all subsequent insertion loss measurements are
made relative to the insertion loss of the cable used as the THRU standard.
For example, if you use a cable with 1 dB of loss, then after Normalization,
the display will show 0 dB of loss with this cable in place. Therefore, for
highest accuracy, when measuring the DUT also attach the cable that was
used in the normalization cal.
— Press Finish.
— Connect the DUT.
CAL ON U
correcting ONLY the appropriate measurement. For example, when an Open
Response Cal on Port 2 is performed,
measurement only.

View Cal

From the Choose Calibration screen (see
page
Keysight N9938-90003 User's Guide
Response Cal
— Select either
Open Response
calibrated. The availability of this calibration on Port 2 may require an
option.
— Connect an OPEN or SHORT standard to the specified port and press
Measure
— Select
Normalization
— Connect a short, high-quality, phase stable cable between the
FieldFox port 1 and port 2 connectors, then press
is shown on the screen when a User Cal (Ex: Response Cal) is
89), press View Cal to see the following screen:
or
Short Response
for the port to be
Measure
CAL ON U
is shown for an S22
"How to Perform a Calibration" on
97