SA (Spectrum Analyzer) Mode (Option 233–Mixed Analyzers)
Spectrum Emission Mask (SEM) — (SA Mode Only)
5. Setting the Detectors
Keysight N9938-90003 User's Guide
Abs AND Rel
failure to generate a fail message.
The Trace menu shows that two Detector controls are available, one for
the Reference Channel (default Auto is Average to facilitate the Power
Integration), and another for all the Offset segments (default Auto is PEAK,
to facilitate finding PEAK signal energy when comparing against Limits).
There is only 1 Trace available, and like the other Channel Measurements,
it is set to average.
Press Trace 6 then
State
- sets the measurement state. Refer to
(SA Mode)" on page
Default All
- Sets the SEM settings to default
Chan Detector
- Sets the Reference Channel detector (default Auto
is Average to facilitate the Power Integration). See also
Method" on page
Offset Detector
Peak, to facilitate finding peak signal energy when comparing
against the Limits). See also
Record Playback
page
463.
- Requires a relative and an absolute data
141.
151.
- Sets all of the Offset segments (default Auto is
"Detection Method" on page
- Refer to
"Record/Playback (SA Option)" on
"Trace Display States
"Detection
151.
187